How Does An NDIR CO2 Sensor Work?

how-does-an-ndir-co2-sensor-work

Share This Post

NDIR CO2 sensors detect and monitor CO2 in the air based on the absorption of infrared (IR) light at a given wavelength. An IR lamp directs light waves through a tube towards a filter. The amount of IR light that passes the filter is measured by an IR light detector. 

Carbon dioxide (CO2) is an odorless and colorless gas formed during combustion, respiration, and organic decomposition. Measuring CO2 is important in a variety of industries and applications from measuring indoor air quality to industrial beer brewing. 

A non-dispersive infrared (NDIR) sensor is typically used to detect and monitor CO2 in the air, also known as NDIR detection. 

How Do NDIR CO2 Sensors Work?

Every element on our planet can absorb certain types of light. For example, grass reflects green light and absorbs all the others (red, orange, yellow, blue, indigo, and violet) which is why our eyes see that grass is green. NDIR sensors are able to pinpoint atoms and molecules that we cannot see, making them extremely useful in detecting CO2 levels. 

As a whole, NDIR CO2 sensors detect and monitor CO2 (usually in the air) based on the absorption of infrared (IR) light at a given wavelength.

An NDIR CO2 sensor consists of an IR source, a light tube, a filter, and an IR light detector. The IR lamp directs light waves through the light tube (filled with air) where the gas is pumped or diffused towards the filter. 

The amount of IR light that passes the filter is measured by the IR light detector which is then converted into an electrical output measurement. 

When detecting CO2, the most commonly used wavelength is 4.26 µm (microns). This wavelength is not absorbed by other gases or water vapor, therefore cross-sensitivities and interference by humidity are reduced. I suppose you could say it is much like a fingerprint. Just like a fingerprint, the wavelength to detect the CO2 molecule in the NDIR sensor is unique. 

When using an NDIR sensor, the more light that is absorbed, the higher the CO2 concentration measured will be. This is because more gas molecules result in a lower output signal. 

The performance of an NDIR CO2 sensor depends on the light emitter and infrared sensor that is installed, which comes in two types seen below. 

Infrared (IR) Light Emitter

Infrared light emitters can be split into two types: optical and thermal. 

Optical emitters convert a current into light using electrons and holes inside a semiconductor. Recombination occurs when a voltage is applied to a light-emitting diode (LED), moving the electrons and holes inside a semiconductor. 

LED optical light emitters respond very fast compared to thermal emitters, therefore they are widely used in NDIR CO2 sensors. They also consume less power and do not overheat because they convert a current directly into light. 

Optical Emitter

Thermal emitters emit light by producing a current, heating the object. A filament in a lamp receives a current, generating heat. This heats the filament and emits light due to an increase in temperature. This type is an old-fashioned IR light emitter, yet it is still very useful for CO2 gas sensors. 

Thermal Emitter

Infrared (IR) Sensor

Infrared light sensors can also be split into optical and thermal types. 

Optical IR sensors (photodiode) use photovoltaic power in a semiconductor to convert light into a current. 

A thermal IR sensor detects a voltage (or polarization) when a temperature change occurs from a warming object. Thermal IR sensors include thermopile and pyroelectric sensors. 

Applications For NDIR CO2 Sensors

  • Indoor air quality
  • Cycle regulation in self-cleaning ovens
  • Fuel gas emissions
  • Brewing beer & other beverages
  • Greenhouse farming
  • Warning signals in hazardous zones
  • Gas leak detection sensors
  • Landfill gas monitoring 
  • Alcohol breathalyzers
  • Monitoring patients undergoing anesthesia

Summing Up NDIR CO2 Sensors

An NDIR sensor detects and monitors CO2 in air based on the absorption of infrared (IR) light at a given wavelength.

Air enters the NDIR sensor which activates a light set at a specific wavelength (4.26 µm) measuring the CO2 level. Once the light is activated, any CO2 inside the sensor will be recorded. The light absorption rate depends on how much CO2 is present in the air; the more CO2, the more light will be absorbed. 

If you would like to learn more about CO2 sensors, or what other testing kits we have to offer, do not hesitate to contact our world-class team at Atlas Scientific.

CO2 Sensor & Scrubber

Subscribe To Our Newsletter

Get product updates and learn from the best!

More To Explore

how-to-calibrate-ezo-complete-tmp-using-atlas-desktop-software
Blog

How to calibrate the EZO Complete-TMP using the Atlas Desktop software

The Atlas Scientific line of EZO Complete circuits makes building a custom monitoring system easy, but if you don’t calibrate your sensors, then what good are your readings; And can you trust them? The accuracy of your readings is directly related to the quality of your calibration. Calibration is not difficult, and a little bit

how-to-calibrate-ezo-complete-dissolved-oxygen-using-atlas-desktop-software
Blog

How to calibrate the EZO Complete-Dissolved Oxygen using the Atlas Desktop software

The Atlas Scientific line of EZO Complete circuits makes building a custom monitoring system easy, but if you don’t calibrate your sensors, then what good are your readings; And can you trust them? The accuracy of your readings is directly related to the quality of your calibration. Calibration is not difficult, and a little bit

Want to learn more about our products?

Scroll to Top

To track your order please enter your Order ID in the box below and press the "Track" button. This was given to you on your receipt and in the confirmation email you should have received.