What Is pH Slope, And Why Is It Important?

Share This Post

A pH slope is a linear correlation between the voltage reading of a pH sensor reading and the pH value displayed. The pH slope is what determines the amount the voltage reading must change to change the pH value by one. It is important to calculate the pH slope because it is a direct indication of how the change in voltage correlates with a change in the pH value. Typically, the voltage will change by 59.16 mV per one-unit change in pH value. 

Measuring pH is essential in a wide range of applications, from testing the pH in beer to monitoring the pH of soil. When working with pH measurements, you can adjust the pH sensor slope via calibration. This is important when testing the pH in applications to ensure the pH sensor is functioning accurately, even if conditions are not ideal.

What Is A pH Slope?

A pH sensor is a device that measures the acidity or alkalinity of a solution. The pH value is measured using a probe that generates a small voltage that is proportional to the pH value of the solution. The pH sensor’s slope is the linear correlation between the voltage reading of the sensor and the pH value. The slope is a numerical indication of how the change in voltage correlates to a change in pH. The pH slope is important because it helps calibrate the sensor to ensure accurate and precise pH measurements.

The theoretical slope value is 59.16 mV at 25 °C, which means that for every change of 59.16 mV in voltage, whether it be a positive (+) or minus (-) change, the pH value will change by one pH unit. 

In ideal conditions, the voltage will change by 59.16 mV for every unit of change in the pH value. However, in real-world conditions, the slope value may vary due to various factors, such as temperature, aging, and contamination.

To calibrate the pH sensor, the slope value needs to be determined. The slope value can be calculated by measuring the voltage generated by the pH sensor in two different buffer solutions with known pH values. The difference in voltage between the two buffer solutions is divided by the difference in pH values to obtain the slope value. The slope value can then be used to adjust the pH sensor’s output voltage to ensure accurate and precise pH measurements.

What Is The pH Range For Slope?

The ideal pH range for slope is 95% to 105%. However, a pH slope between 85% and 105% is also acceptable. If the pH slope is outside this range, it may be an indication that your pH probe needs replacing. Therefore, the pH slope range varies depending on how well your pH probe is working properly.

Why Are pH Slopes Important?

The pH slope is an essential factor in pH measurement because it affects the accuracy and precision of the measurement. If the pH slope is not calibrated correctly, the pH readings may be inaccurate, leading to errors in the analysis results. As mentioned, the pH slope is affected by various factors, such as temperature, aging, and contamination. Therefore, it is essential to calibrate the pH sensor regularly to ensure accurate and precise pH measurements.

The pH slope also affects the sensitivity of the pH sensor. A steep slope means that the pH sensor is highly sensitive to changes in pH, while a shallow slope means that the pH sensor is less sensitive to changes in pH. Therefore, a steep slope is desirable for pH sensors used in applications where small changes in pH need to be detected.

Another reason why pH slopes are important is that they can be used to diagnose problems with the pH sensor. If the pH slope value is not within the expected range, it may indicate that the pH sensor is not functioning correctly, and further investigation is needed to identify and rectify the problem.

What Is The pH Range For Slope?

The pH range for slope is the pH range over which the pH sensor’s slope is valid. The pH range for slope depends on the type of pH sensor and the buffer solutions used to calibrate the sensor. For most pH sensors, the pH range for slope is between pH 4 and pH 10. Outside this range, the slope value may deviate from the theoretical value, leading to inaccurate pH measurements.

The pH range for slope is affected by various factors, such as temperature, aging, and contamination. Therefore, it is essential to calibrate the pH sensor regularly to ensure that the pH range for the slope is valid.

Calculating pH Slope

To calculate the pH sensor’s slope, two buffer solutions with known pH values are required. The buffer solutions should be chosen based on the pH range of the sample being measured. For example, if the sample has a pH of 7, buffer solutions with pH values of 4 and 10 can be used.

The pH sensor is immersed in the first buffer solution, and the voltage generated by the sensor is measured. The pH sensor is then immersed in the second buffer solution, and the voltage generated by the sensor is measured again. The difference in voltage between the two buffer solutions is divided by the difference in pH values to obtain the slope value.

To calculate the pH slope, follow the steps below:

  1. Using a pH 7.0 buffer, measure the mV value – this gives you the electrode offset
  2. Next, using a second buffer (4.0 or 10.0), measure the mV value
  3. To work out the pH electrode slope, calculate the difference between the mV value of the first and second buffer
  4. Divide this value by the difference of pH units between the two buffers For example, if you are using a 7.0 and 10.0 buffer, the difference is 3. 
  5. Now you want to convert this number to the electrode slope percentage. Divide the electrode slope (59.15 mV) by the maximum slope.
  6. Finally, multiply the number by 100.

Let’s use an example. 

If the pH electrode generated a -10 mV in pH 7.0 and +165 mV in a pH 4.0 buffer, the value would be 170 mV.

  • 165 mV – (-10 mV) = 170 mV
  • 170 mV/3 = 58.33 mV/pH unit  
  • 58.33/59.16 x 100 = 98.60%

Summing Up, pH Slopes

In conclusion, pH slope is an essential factor in pH measurement, affecting the accuracy, precision, and sensitivity of the pH sensor. The pH slope needs to be calibrated regularly to ensure accurate and precise pH measurements. 

The pH range for slope depends on the type of pH sensor and the buffer solutions used to calibrate the sensor. Calculating the pH sensor’s slope requires two buffer solutions with known pH values and can be done using a simple equation.

If you have any questions regarding pH slopes or are unsure which pH probe will best suit the application you are working with, do not hesitate to contact the world-class team at Atlas Scientific. 

pH Probes

Subscribe To Our Newsletter

Get product updates and learn from the best!

More To Explore

how-to-calibrate-ezo-complete-tmp-using-atlas-desktop-software
Blog

How to calibrate the EZO Complete-TMP using the Atlas Desktop software

The Atlas Scientific line of EZO Complete circuits makes building a custom monitoring system easy, but if you don’t calibrate your sensors, then what good are your readings; And can you trust them? The accuracy of your readings is directly related to the quality of your calibration. Calibration is not difficult, and a little bit

how-to-calibrate-ezo-complete-dissolved-oxygen-using-atlas-desktop-software
Blog

How to calibrate the EZO Complete-Dissolved Oxygen using the Atlas Desktop software

The Atlas Scientific line of EZO Complete circuits makes building a custom monitoring system easy, but if you don’t calibrate your sensors, then what good are your readings; And can you trust them? The accuracy of your readings is directly related to the quality of your calibration. Calibration is not difficult, and a little bit

Want to learn more about our products?

Scroll to Top

To track your order please enter your Order ID in the box below and press the "Track" button. This was given to you on your receipt and in the confirmation email you should have received.